Monika Jenko
Institute of Metals and Technology
Title: Premature fracture of a Modular Femoral Neck after Total Hip Arthroplasty: Comparison of Two Different Stem-Neck Alloys
Biography:
Dr Monika Jenko accomplished her PhD studies of material science at the University of Ljubljana. She was 11 years director of the public research Institute of Metals and Technologies (IMT) and initiator that institute has joined to Jozef Stefan International postgraduate School with Advanced Metallic Materials in the frame of the Nanoscience and nanotechnology study program. She introduced the surface analysis in Slovenia into metallurgy as well as the integrated electron spectroscopy techniques. She has worked in different national and international projects in the field of Material Science, Applied Surface Science, Surface Engineering, Nanoscience and last ten years she is very active in biocompatible materials . She is author of several papers published in reputed journals and is a member of different international working groups
Abstract:
Statement of the Problem: An alarmingly large number of femoral modular stems of titanium alloy neck suffer from premature fractures. The cobalt-chromium neck was introduced for its better wear and mechanical properties, although it fails even earlier. Until now there were no direct comparisons between two different femoral modular necks for the same type of stem implant. Experts from orthopaedic surgery and metallurgy have combined their research efforts to compare the femoral modular stem-necks of similar titanium and dissimilar titanium/cobalt-chromium alloy systems from the clinical and metallurgical points of view.
Orientation: Two premature fractured modular neck Hip endprostheses of similar and dissimilar metals systems were investigated. Multivariate analyses were performed to assess the differences in the fretting, corrosion and fatigue of similar alloys and dissimilar alloys modular joints. Patient’s demographic information was collected from medical records.
Findings: Similar stem-neck metals of hip endoprostheses failed due to continuous dynamic micro-motions, crevice corrosion, fretting and fatigue, as well as improper neck dimensions. The mechanism of titanium alloy modular neck fracture is shown in Figure 1. The dissimilar metals stem-neck fractured due to continuous dynamic micro-motions, selective leaching of cobalt ions from the cobalt-chromium-molybdenum alloy and formation of titan-chromium-molybdenum interfacial phase and severe crevice corrosion. The mechanism of cobalt-chromium-molybdenum alloy modular neck fracture is shown in Figure 2 The dissimilar joints suffered more corrosion than the similar joints due to additional galvanic corrosion.
Conclusions: Fretting, corrosion and fatigue occurred on both neck-stem retrievals of the similar and dissimilar metals. Both metal-alloy systems used in this application are known to be highly corrosion resistant and giving rise to the bio-tribo-corrosion processes needs to be understood and characterized so that appropriate changes in design and materials can be upgrade.